7.4
Cycle setting
TEMPERATURE from 20 to 65°C
To set the cleaning temperature: press the TEMP button; by doing so, the led can be
scrolled through the setting BAR; stop at the desired temperature.
TIME from 1 to 30 minutes
To set the cleaning time: press the TIME button; by doing so, the led can be scrolled
through the setting BAR; stop at the desired cleaning time.
DEGAS
Press the DEGAS button: if the corresponding LED lights up, the function is set
SWEEP
Press the SWEEP button: if the corresponding LED lights up, the function is set
At the end of the cleaning cycle, settings are stored and re-proposed for the next
cycle.
DEGAS and SWEEP
The DEGAS and SWEEP functions can be enabled and disabled by pressing the
appropriate buttons.
These functions can be set before starting a cycle or enabled during cleaning.
The duration is determined by the initial setting of the cleaning time.
The DEGAS and SWEEP functions are mutually exclusive: if DEGAS is set and
SWEEP is pressed, the DEGAS function is automatically disabled while the SWEEP
one is enabled.
AUTOMATIC FUNCTIONS
They are enabled by pressing and holding a specific button for longer than 6 sec.
DEGAS: By pressing and holding the corresponding button for a time ≥ 6 sec, the
device automatically performs a 5-minute DEGASSING cycle; it is not necessary to
press the START button. To stop, press the STOP button before the 5 minutes have
elapsed.
CYCLE START ONLY IN THE VERSIONS WITH HEATING
It is possible to start a cleaning cycle automatically after the device has reached the
set temperature.
PRESS AND HOLD the START button for a time ≥ 6 sec. The device starts
operating; the operation LED starts flashing; the cleaning liquid starts heating up.
When the set temperature is reached, the cleaning cycle will start and run for the
set time.
(VERSION WITH HEATING)
ENGLISH
17