Original Operation Manual
Classification of equipment
Type of protection against electric shock :
- Class II equipment :
Degree of protection against electric shock :
- Type B applied part :
Method of sterilization or disinfection recommended by the manufacture :
- See 7. STERILIZATION
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrousoxide :
EQUIPMENT not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrousoxide.
Mode of operation :
- Continuous operation
Cautions for handling and operation
Read these safety cautions thoroughly before use and operate the product properly.
These instructions will show you how to operate the product safely and prevent danger to you or others. They are
classified by the degree and/or severity of danger. All contents relating to safety should be observed.
Classification
W A R N I N G
Provides an instruction where personal injury or physical damage may occur.
C A U T I O N
Provides an instruction where minor to medium injury or physical damage may occur.
N O T I C E
Provides an instruction that should be observed for safety reasons.
· TO PREVENT ELECTRIC SHOCK Do not unplug the power cord with wet hands.
· TO PREVENT ELECTRIC SHOCK Be sure to prevent water on the Control Box.
· TO PREVENT ELECTRIC SHOCK Do not unplug the power cord roughly.
· TO PREVENT ELECTRIC SHOCK Use an electrical outlet that is grounded.
· If you feel any abnormality such as vibration, heat generation, abnormal noise, etc., prior or during use of the unit, stop
using it immediately.
· Use the fuse of specified rating. [AC20V: T.6A 250V (Ref No.FU00, FU0), AC230V: T800mA 250V (Ref No. FU00,
FU0)
· When operating this system always consider the safety of the patient.
· The product is designed only for clinical dental use by qualified personnel.
· Every connection must be firmly connected. If it is not firmly connected, it is because an air/water leak or LED does not on.
· Do not allow any impact on to the Product. Do not drop the Product. It may cause a malfunction.
· This system can be used with ISO 968, type B or type C type hose connector.
· The system functions normally in the environment where the temperature is at 0-40°C (32-04°F), humidity at 0-85%
RH, atmospheric pressure at 700-060hPa, and no moisture condensation in the Control Unit. Operating at outside of these
limits may cause malfunction.
· When you install Control Box or Motor, DO NOT bend or twist the tubing or the cord. No unnatural force is needed to do the
installation.
· Do not use contaminated air (by dust or moisture). If the air contains water or dust, it might cause a malfunction or
overheat.
· Do not connect NLX nano micromotor to Ti-Max NL400. It will prevent the LED from working properly.
Degree and severity of danger or damage