5.6. CONTINUITY TEST AND DIODE TEST

Before attempting any resistance measurement, remove power from the
circuit under test and discharge all capacitors, if present.
Fig. 8: Use of the instrument for continuity test and diode test
1. Select the position 
2. Press the MODE key until the symbol " " is displayed to activate continuity test.
3. Insert the red cable into input terminal VHz%
terminal COM and carry out the continuity test of the object to be measured (see Fig.
8– left side). An buzzer sounds when the measured value of resistance is lower than
30
4. Press the MODE key to select diode test. The symbol "
5. Connect the red lead to the anode of the diode and the black lead to the cathode in
case direct polarization measurement is carried out (see Fig. 8 – right side). Invert the
position of the leads in case reverse polarization measurement is carried out.
6. Values on the display between 0.4V and 0.7V (direct) and "O.L" (reverse) indicate
correct connection. A value "0mV" indicates that the device is short-circuited, while
"O.L" in both directions indicated an interrupted device.
CAUTION
EN – 13
and the black cable into input
" appears on the display.