• Do not sterilize by ultraviolet ray. The handpiece could discolor.
• If autoclaved with other instruments stained with chemical solution, it could strip the plating
CAUTION
and make the surface black.
• Do not autoclave any parts (the Control Unit, AC Adaptor, Foot Control, Handpiece Cord,
O-Ring). Other than those that can be subjected to autoclave sterilization. Perform alcohol
disinfection to the Control Unit, AC Adaptor, Foot Control, Handpiece Cord including after
every patient.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
• If the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the
dry cycle.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and
sunlight. The air should be free from dust, salt and sulphur.
• Do not touch the product immediately after autoclaving as it will be very hot and must remain
in a sterile condition.
13
Troubleshooting
When trouble is found, please check the followings prior to consulting your dealer.
Problem
Probable Cause
The Power
Indicator
does not turn
on,even if the
power is ON
The Tip does
not generate
No / Poor
vibration,
vibration
in spite of
depressing the
Foot Control
The Tip is bent
or broken
The Tip falls off
Noise from the
handpiece
The AC Adaptor or the DC Plug
is disconnected
The Tip is not tightened firmly
Worn Tip
Power has not been correctly
adjusted for the Tip
The Foot Control is disconnected Connect the Foot Control correctly
Failure of vibrator in the
handpiece
Failure of internal components
of the Foot Control
Power has not been properly
—
adjusted for the Tip
—
The Tip is not tightened firmly
Power has not been properly
adjusted for the Tip
—
The Tip is not tightened firmly
Failure of vibration in the
handpiece or the Control Unit
Cause
Correctly insert the AC Adaptor or the DC
Plug
Tighten the Tip until the Tip Wrench clicks
Replace the Tip
Adjust the power level following the
Power Guide or Tip case label. Do not
exceed the power limit for the tip.
Contact dealer*
Contact dealer*
Adjust the power level following the
Power Guide or Tip case label. Do not
exceed the power limit for the tip.
Tighten the Tip until the Tip Wrench clicks
Adjust the power level following the
Power Guide or Tip case label. Do not
exceed the power limit for the tip.
Tighten the Tip until the Tip Wrench clicks
Contact dealer*
14
Solution