NSK Varios 370 Mode D'emploi page 4

Masquer les pouces Voir aussi pour Varios 370:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
The surface of the Control Unit may feel warm after prolonged use. Turn the Power/Volume Knob to OFF
and cool the device.
Users are responsible for operational control, maintenance and inspection.
Clean/sterilize the product immediately after using it. Then store it. Leaving it non-sterile might lead to failure.
When you have not used the product for long time and use it again, check the operation before use.
Eye damage may result if the LED is stared directly into. Do not look into or turn it to the eyes of the patient.
When abnormalities are found with a Control Unit and /or an AC adaptor, pull the AC adaptor from AC Outlet
immediately.
Do not turn the Power/Volume Knob to ON or OFF needlessly.
No special training is required for this device.
Applied parts for patient and/or operator are/is Tip and Handpiece.
When using a file as a tip for root canal treatment (for enlarging, cleaning, etc.), use it after confirming its
characteristics and areas where it can be applied.
Surface temperature of tip shall be more than 50˚C without using coolant water. To avoid this event, be sure
to use coolant water.
* Operation Principle
A sinusoidal electrical signal, at ultrasonic frequency ( f > 20kHz ), is delivered by the generator. This
signal is applied to the 'piezoelectric ceramic' located inside the transducer. Piezoelectric ceramic
converts this signal into mechanical vibrations. These vibrations are at the same ultrasonic frequency as
the electrical signal. The mechanical vibrations are propagated towards the distal end of the transducer.
The "TIP" insert, which is attached at the distal end of the transducer, vibrates at ultrasonic frequencies
and makes it possible to achieve the aimed purpose.
CAUTION
3

Publicité

Table des Matières
loading

Table des Matières