5) Remove the probe from the control unit then insert the Tester directly into the
probe connector. Check that the number display on the LCD panel is within
the range of "0.4 and 0.6".
●
CAUTION
●
●
●
8
Audio Alarm Volume Control
Alarm audio volume can be adjusted to "OFF", "Low ( ) ", "Medium (
(
) ".
1) Press the Alarm key.
2) The Alarm Mark on the LCD panel and the sound volume change.
3) Each time the Key is pressed the sound volume changes.
●
CAUTION
●
NOTICE
12
If the number is not displayed correctly, check the following.
- That the tester, probe, and file clip are securely connected.
- That the tester and file clip are not wet, or skin is not in contact
with the electrode of the tester.
If the number display is outside the range of "0.4 and 0.6" or the
value is not correctly indicated in step 5, the control unit may have
an abnormality. Contact your Authorized NSK Dealer.
If the number display is outside the range of "0.4 and 0.6" in step
3 and 4 even though it operates correctly in step 5, there may be a
fault in any cord. Contact your Authorized NSK Dealer.
After operation check, be sure to autoclave the file clip and lip hook
for next use.
Do not press the Alarm Key too strongly as it may be damaged.
The last setting is stored when the control unit is switched OFF.
) " and "High