CARE AND MAINTENANCE: How to replace the in-use cover and the fuse
To protect the balance housing against soiling, damage and corrosive chemicals, the balance is delivered with the in-use cover already in place. This cover is made
of stiff plastic so that it cannot be folded.
The in-use cover does in no way hinder your working with the balance.
A back-up fuse is built into the fuse holder, in case your microfuse becomes defective.
If the microfuse has to be replaced repeatedly without apparent reason, call METTLER TOLEDO Service.
Removing the in-use cover
– Remove weighing pan and pan support.
– Turn retaining ring (25) counterclockwise until
it is released and can be lifted off.
– Remove in-use cover (26).
– Install pan support and weighing pan.
(The retaining ring should only be installed when
the in-use cover is in place).
Installing the in-use cover
– Remove weighing pan and pan support.
– Place in-use cover (26) on balance.
– Position and engage retaining ring (25), then
turn clockwise to stop position.
– Install pan support and weighing pan.
Replacing the microfuse
– Disconnect power cable.
– Using a screwdriver, remove fuse holder (4).
– Remove defective fuse and insert new fuse
(back-up fuse from fuse holder).
Fuse rating:
125 mA, slow-acting, for 95/120 V
63 mA, slow-acting, for 190/240 V
– Insert fuse holder.
– Plug in power cable.