Fault – Cause – Remedy
Fault – Cause – Remedy
Fault
Apparatus does not commence inspira-
tion when triggered.
No aerosol from nebulizer.
Inhalog 2 plus will not start.
Inhalog 1 plus
Green LED »O
« does not light up.
2
18
Cause
Leak or defect in hose connection or
nebulizer.
Medical air supply not connected to Inha-
log 1 plus or supply pressure too low.
Trigger faulty.
Filling level in medicament nebulizer too
low.
Nebulizer soiled.
Nebulizer hose not connected to appara-
tus or medicament nebulizer.
Apparatus faulty.
No electric power.
Compressor in Inhalog 2 plus faulty.
O
supply not connected or supply pres-
2
sure too low.
Remedy
Plug hoses in securely, see page 6 et
seq., replace defective hoses; check
nebulizer, assemble correctly, replace
defective parts.
If used:
check disposable patient system and
replace if necessary.
Connect medical gas hose, check supply
pressure. Supply pressure must be 3.0
to 6.0 bar.
Call DrägerService.
Refill medication.
Thoroughly clean nebulizer.
Plug in nebulizer hose.
Call DrägerService.
Check that mains connector is plugged
in, check power supply. Check fuses and
replace if necessary, see page 19.
Call DrägerService.
Connect O
hose, check supply pres-
2
sure.