7. Troubleshooting
Before calling technical support, please check that your problem cannot be
solved easily by following the guidelines below. In this chart, you will find the
most common problems and their solutions.
PROBLEMS
CHECK
Check that the power cord is correctly connected.
Check that the switch is on "I" (ON).
Unit does not start
Check the fuses and the mains voltage are ok.
Check that the door closes well. If needed clean the
blending chamber.
The screen indicates
This alarm is activated when an overvoltage is detected in the
"Err I"
motor. Increase the gap between the door and the paddles.
The screen indicates
This alarm is activated when the motor is blocked. Check that
"Err 2"
nothing prevents the mechanical operation of the device.
The blending time is
Set the timer (Time) on the appropriate blending time.
not convenient
The blending speed
Set the speed (Speed) on the appropriate blending speed.
is not convenient
Thermo Scientific
7
EN
Homogenizer
PROBLEMS
The blending bags
leak during the
blending process
The blending bags
leak during the
blending process for
hard or heavy
samples
(wood, cloths...)
The sample spits
over the door
18
|
Thermo Scientific
CHECK
Check the quality of your bags.
You may increase the space between the door and the
paddles of 5 mm and/or decrease the speed of blending.
Check the state of the seal on top of the blending chamber. If
needed, turn it a quarter turn. (Contact Customer support)
EN
Troubleshooting
|
Chapter 7
Homogenizer
|
19