Technical Data, Options, Optional Equipment; Technical Data - Mettler Toledo AL Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Technical data, options, optional equipment

22
6 Technical data, options, optional equipment

6.1 Technical data

Standard equipment of L/L-S balances
AC adapter to national standard as per list in Section 6.4.
Balance power input 6-14,5VAC, 50/60Hz, 4VA or
7-20VDC, 4W
Draft shield (on models with 0.1 / 1 mg resolution)
All models can weigh below balance
External adjustment weight with AL models
Materials
Housing base:
standard construction: die-cast aluminum, painted;
compact construction: plastic (ABS/PC)
Top housing: plastic (ABS/PC)
Weighing pan: 18/10 chromium-nickel steel
Batteries, disposable and rechargeable (compact
construction)
Disposable: 4 x AA (LR6) 1.5 V alkali-manganese, typical
20 h (with 2.9 Ah capacity)
Internal battery charger "AccuModule" (optional):
4 NiMH, typical 11 h/charging time 5 h (with 1.5 Ah
capacity)
Readability
Max. capacity
Repeatability (sd)
Linearity
Sensitivity temperature drift
(10 °C ... 30 °C)
Settling time, typical
Adjustment weight external
Construction
External dimensions of balance
(W/D/H) in mm
External dimensions of packaging
(W/D/H) in mm
Weighing pan
Usable heigh of draft shield
Net weight (with packaging) kg
Level indicator
Number of leveling screws
Protection
Ambient conditions
The technical data are valid under the following ambient
conditions:
Operability is assured at ambient temperatures between 5 and
40 ºC.
AL54
AL104
0.0001 g
0.0001 g
51 g
110 g
0.0001 g
0.0001 g
0.0002 g
0.0002 g
2.5 ppm/°C
2.5 ppm/°C
4 s
4 s
50 g
100 g
Standard
238x335x364
520x385x555
ø 90 mm
220 mm
5.8 (8.2)
yes
yes
2
2
Protected against dust and water
Pollution degree: 2
Installation category: class II
EMC: see declaration of conformity
Ambient temperature
Relative humidity
AL204
0.0001 g
210 g
0.0001 g
0.0003 g
2.5 ppm/°C
4 s
200 g
yes
2
10 °C ... 30 °C
10 % to 80 % at 31 °C, linear
decreasing to 50 % at 40 °C
noncondensing

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

PlPl-s

Table des Matières