12. Notes on electromagnetic compatibility
Guidelines and manufacturer's declaration - electromagnetic immunity
The 7835 series scale is designed for operation in the ELECTRICAL MAGNETIC ENVIRONMENT specified below. The
customer or the user of the 7835 scale should ensure that it is used in such an environment.
Interference immunity
test
Static electricity
discharge (ESD) accor-
ding to IEC 61000-4-2
Fast transient electrical
disturbances/bursts ac-
cording to IEC 61000-4-5
Surges according to IEC
61000-4-5
Voltage dips, short inter-
ruptions and supply
voltage fluctuations
according to IEC
61000-4-11
Magnetic field at the
supply frequency (50/60
Hz) according to IEC
61000-4-8
NOTE: LT is the mains alternating voltage before the test level is applied
54
IEC 60601 test level
Compliance level
+ 6 kV contact
+ 6 kV contact
discharge (indirect)
discharge
+ 8 kV
+ 8 kV
Air discharge
Air discharge
+ 2 kV
+ 2 kV
for power cables
for power cables
+ 1 kV
+ 1 kV
for input and output
for input and output
lines
lines
+ 1 kV
+ 1 kV
Voltage outer
Voltage outer
conductor - outer
conductor - outer
conductor
conductor
± 1 kV
Not applicable
Voltage outer
conductor - earth
< 5% UT
< 5% UT
for ½ period
for ½ period
(> 95 % collapse)
(> 95 % collapse)
40% UT
40% UT
for 5 period
for 5 period
(60 % collapse)
(60 % collapse)
70% UT
70% UT
for 25 period
for 25 period
(30 % collapse)
(30 % collapse)
< 5% UT
< 5% UT
for 5 s
for 5 s
(> 95 % collapse)
(> 95 % collapse)
3 A/m
3 A/m
Electromagnetic environment -
Guidelines
Floors should be made of wood or
concrete or be covered with ceramic
tiles. If the floor is covered with
synthetic material, the relative
humidity must be at least 30%.
The quality of the supply voltage
should be that of a typical business or
hospital environment.
The quality of the supply voltage
should be that of a typical business or
hospital environment.
The quality of the supply voltage
should be that of a typical business or
hospital environment.
If the user of the 7752 scale requires
continued FUNCTION even in the
event of power supply interruptions, it
is recommended that the 7752 scale
be powered from an uninterruptible
power supply.
Magnetic fields at the mains
frequency should correspond to the
typical values found in the business
and hospital environment.