2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
8. Using a Mains Adapter
You can operate this device using the Microlife mains adapter
(DC 6V, 600 mA).
Only use the Microlife mains adapter available as an original
accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 5 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
9. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat the
measurement.*
BP A3 Easy
Error
Description Potential cause and remedy
«ERR 2»
Error signal During the measurement, error signals
were detected by the cuff, caused for
AS
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«ERR 3»
No pressure
An adequate pressure cannot be gener-
in the cuff
ated in the cuff. A leak may have
AP
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«ERR 6» MAM Mode There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure
299 mmHg) OR the pulse is too high
too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
10. Safety, Care, Accuracy Test and Disposal
Safety and protection
• Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
• This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
EN
5