How to inhale with the OMRON MicroAIR U100
Note: If high-viscosity medication is used, nebulization may be
reduced.
If excessive medication collects on the mesh, nebulization may
stop. In that case, turn off the power and remove the mask
adapter. Absorb the medication by means of a lint-free cloth.
Note: Do not poke the mesh with a cotton swab or a pin; the mesh
may be permanently damaged.
After inhalation always switch off the nebulizer by pressing the ON/OFF ( ) button.
Note: The device automatically turns off after 30 minutes.
18
Absorb the medication.