Start up and shut down

4.3 Checking the pressures

CAUTION
Compressed air connections on device.
Dirty and moist compressed air causes premature bearing wear.
▶ Ensure dry, clean and uncontaminated compressed air according
to ISO 7494-2.
Insert the test gauge (see spare parts catalog) between the
QC02 LED and the medical device,
and check the following pressures:
- Drive pressure: 2.8 bar (40 psi)
- Return air: < 0.5 bar (7 psi)
- Water: 0.8 to 2.5 bar (12 to 36 psi)