4.2.4 XSE balance settings for Antistatic Kit
The arrow buttons can be used to page forward or back to a menu page.
1
Switch on the balance [ ].
2
Press [ ].
3
Choose Smart & ErgoSens and tap the button [Define].
A selection window appears.
4
Choose ErgoSens 1 (Aux1) or ErgoSens 2 (Aux2) and tap on the associated button.
A selection window appears.
5
Tap in the scroll bar to select [AntiStatic Kit].
6
To save the settings, tap [OK].
7
Choose Function Keys and tap the button [Define].
A selection window appears.
8
Tap [Ionizer].
9
To save the settings, tap [OK].
The function key [Ionizer] is displayed in the application at the bottom of the display.
Indicator light turns on and Status-Icon [ ] appears in the display when the Ionizer is working.
5 Operation
The ionizer is controlled by the balance. It should be activated during loading the balance and deacti
vated during weighing.
Activating with XPE
With XPE balances the ionizer is controlled by the automatic door.
Activating with XSE
–
To activate the ionizer, tap function key [Ionizer].
The indicator light on the ionizer turns on.
–
To deactivate the ionizer, tap again function key [Ionizer].
The indicator light turns off.
If the indicator light doesn't turn on:
1
Check cable connections and settings.
2
If the indicator light remains off, contact your local METTLER TOLEDO service agency.
Further information, see balance Operating Instructions.
8
Operation
CAUTION
Risk of ignition or sparking
The ionizer must NOT be used inside an inert gas atmosphere or with an inert
gas purging system. There is a serious risk of sparks from the ionizer if such
an environment is used.