Automatic Disinfection (External and Internal Surfaces)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
8-6 Lubrication
NSK PANA SPRAY Plus / PANA SPRAY
Apply NSK PANA SPRAY Plus / PANA SPRAY every time after each use and/or before autoclaving.
1) Attach the CORRECT Spray Nozzle onto the spray port nozzle on the can.
2) Insert the Spray Nozzle into the rear of the handpiece. Hold the handpiece and spray
2-3 times for approximately 2-3 seconds each time. Apply lubricant until it expels
from the handpiece head for at least 2 seconds (Fig. 15).
E-Type Spray Nozzle
Handpiece
Fig. 15
17