8-4 Cleaning (Spray Port)
1) Internally clean the Internal Spray Nozzle and/
or the External Spray Nozzle by using Cleaning
Wire. (Fig. 13,14) (X-SG20L / X-SG25L / SG20)
2) Connect the Irrigation Tube to each Spray
Nozzle, and wash the inside with clean water.
(Refer to "6. Spray Nozzle") (X-SG20L /
X-SG25L / X-SG93L / X-SG93 / SG20)
CAUTION
Do not forcibly insert the cleaning wire into the Spray Nozzle. Resultant port damage could cause the spray to be directed
away from the bur/drill, and cause reduction of cooling efficiency.
8-5 Disinfection
Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution such as alcohol.
16
Internal
Cleaning Wire
Spray Nozzle
External Spray Nozzle
Fig. 13
Fig. 14