Programming the Stimulator Stimulation Settings
Delivering Test Stimulation
4
To deliver test stimulation, set the desired stimulation
amplitude on the stimulation bar ( 1 ) (see manual section
Controlling Stimulation Amplitude for more information). Then
press the stimulation button to deliver stimulation ( 2 ). The
stimulation button will light up when Stimulation is on.
1
When stimulation is on, the electrode configuration, ramp
settings, frequency, and pulse width cannot be changed.
Incremental changes can be made to the stimulation amplitude.
3
Save Therapy Settings
After therapy settings have been configured, press "Set
2
Program" ( 3 ) to save the settings to the Stimulator. A prompt
will appear for selecting the maximum amplitude the patient
can adjust their stimulation up to.
To exit, press "Patient Device" to return to the Patient Device
screen ( 4 ).
Note: If "Set Program" is not pressed prior to exiting, the
Stimulator settings will revert back to last saved settings when
the Programming screen is exited. When the Programming
screen is exited, a prompt will appear requesting confirmation to
exit without saving the therapy settings.
Note: The amplitude step size is fixed at 0.10mA.
53