Programming the Stimulator Stimulation Settings
50
1
2
3
5
4
Manually Changing Electrode Configuration
The electrode configuration can be manually set by pressing the
electrode representations ( 3 ) to toggle each electrode between
its possible states. There are 3 possible electrode states:
Cathode (or "-")
Anode (or "+")
Unassigned
Note: When manually adjusting electrode assignment, invalid configurations
may be created. An alert will pop up if an attempt is made to exit this
screen, stimulate, or set the therapy settings ( 4 ) with an invalid electrode
configuration selected.
Typical invalid electrode configurations include:
More than 2 cathodes
2 cathodes that are not adjacent electrodes
More than 1 anode