6 Commissioning
!
!
92
6.4
Device configuration
Note!
The figures show the menu structures with the display and picklist for the different configuration
parameters. Below the figures are tables containing the corresponding function descriptions for the
parameters.
Note!
You can only select switching and event programmes with the software option "7 programmes".
6.4.1
Configuring the inputs
You can assign the inputs when configuring in the "Quick Setup" (see Chapter 6.3).
Digital inputs
Menu structure:
Fig. 30:
Digital input menu structure
Sampling
Bottle
Quantity
Ext. stop
Event
Report
NC contact
NO contact
start aut
The input signal triggers a sample.
The input signal triggers a switch to the next empty bottle.
The input signal is a pulse signal from a volume meter; alternative to an analogue
quantity signal (0/4...20 mA).
The input signal stops any programmes running; the programmes will be continued
when the signal disappears.
The input signal triggers an "event sample". For example, the input signal may be
switched when a limit value is infringed; a separate bottle can be filled in an event
sample.
A message appears in the display showing the date and time (e.g. function error of
flowmeter). You must acknowledge the message; the sample programme is not
interrupted.
Special function for capacitive switch-off (optional)
Special function for capacitive switch-off (optional)
The input signal starts the selected main programme