3 Installation
!
74
Note!
Always transport the cabinet vertically. Do not tilt it!
3.3

Installation conditions

3.3.1

Dimensions

You can find the dimensions of the sampler in the Section »Mechanical construction« on page 124.
3.3.2

Installation site

Fig. 3:
Notes on mounting the device
When mounting the device, please observe the following points:
• Position the device on a level surface.
• Protect the device against additional heat (e.g. radiators).
• Protect the device against mechanical vibration.
• Protect the device against strong magnetic fields.
• Ensure that air can circulate unhindered at the back of the cabinet. Do not position the device
directly against a wall.
- The distance between the wall and the rear side of the cabinet should be at least 100 mm
- The distance between the ceiling and the top side of the cabinet should be at least 450 mm
• Do not position the cabinet directly above the channel of the inflow to the wastewater treatment
plant (sulphurous vapours!).