Sample mode
Menu structure:
event program
Sample >
Distribution >
Parallel sample >
â
Esc
<-'
Fig. 43:
Sample mode menu structure
The event programme contains the same sampling options (time, quantity, ext. signal and flow) as
those of the main and switching programmes. In addition, you can select the function <ONCE>. In
this function, the sampler takes a sample in the event programme once and then returns immedia-
tely to the main programme if the event signal is no longer present.
Distribution
You can reserve separate bottles for the event programme. With the exception of parallel reserve
sampling, the following applies when distributing to bottles:
The first bottle group of a distribution is reserved for the main programmes.
The second bottle group is reserved for the switching programmes.
The last bottle group is reserved for the event programme.
Menu structure:
event program
Sample >
Distribution >
Parallel sample >
â
á
Esc
Fig. 44:
Distribution menu structure
You can define the bottles for the event programme as follows:
no change
There is no bottle change when the system switches to the event programme.
move
The next empty bottle is filled when the system changes to the event programme.
1-9 bottles
When the system switches to the event programme, 1-9 bottles are filled from the last
distribution bottle group. The number of bottles you can reserve for the event pro-
gramme is dependent on the total number of bottles (max. 9 bottles).
Sample
-mode
:
time
time
:
volume
:
shots
:
â
á
Esc
Distribution
-mode
:
time
time
:
switch
: no change
<-'
â
á
Esc
6 Commissioning
time
quantity
00:15 ext. sig.
100ml flow
01 once
<-'
time
quantity
01:00 ext. sig.
<-'
101