122
Sample preservation
The sample bottles are located in the lower section of the sampler. The sample compartment tem-
perature can be set directly at the controls from +2 to +20 °C (factory setting: +4 °C). The current
sample compartment temperature is displayed at the controls and recorded in the internal data log-
ger. The evaporator and defrost heater are packed in the PU insulation behind the inner shell, pro-
tected against corrosion and damage. The compressor and liquefier are located in the upper section
of the sampler.
All parts carrying media (e.g. tap, dosing system, distribution pans) can be easily disassembled and
cleaned without tools. The entire sample compartment is fitted with a seamless plastic inner shell
for easy and effective cleaning.
Dosing unit
Dosing volume
Dosing accuracy
Repeating accuracy
Conveying velocity
Conveying height
Conveying distance
10.2
Power supply
See Section 4.1.5 "Terminal assignment"
230 V AC (+10%; -15%), 50 Hz
110-125 V AC, 50/60 Hz
Fuse protection of the power source max. 10 A
• 2 x scewed cable gland M16
• 2 x screwed cable gland M20
• 2 x screwed cable gland M32
Power supply: e.g. NYY-J, 3-core, max. 2.5 mm
Analogue and signal lines: e.g. LiYY 10 x 0.34 mm
Interface RS485: e.g. LiYCY 2 x 0.25 mm
350 W
Serial interface
– RS485 on terminal board
– RS232 (optional for external data logging), 9-pin SUB-D socket on the front panel
®
PROFIBUS
-DP connection (optional)
With Profibus coupler on top-hat rail in electronics compartment via RS232, baudrate 9600 kBaud.
10.3
Installation conditions
See Section 3.3 "Installation conditions"
20 to 200 ml (20 to 500 ml optional)
4% of the set volume
2%
> 0.5 m/s, to EN 25667
max. 6 m (8 m optional)
max. 30 m
ASP Station 2000
Endress+Hauser