• The tap motor was replaced
• The error message: <Tap calibration> appears in the display.
Carry out the calibration as follows:
Fig. 26:
Tap calibration
1. Switch the device on (see Chapter 6.2).
2. Under < SET -> SERVICE - CALIBRATION>, select the item
< DIST. TAB >.
3. When you select "Start", the tap turns and stops just before
the calibration position.
4. At the controller, keep selecting <1 step> until the arrow on
the front side of the tap is located exactly in the notch in the
middle of the distribution pan (point A). In the picklist, select
the menu option <SAVE>.
5. The tap is calibrated.
5 Operation
87