Storage conditions:
Temperature: -40 – 70 ºC
Humidity: 0 – 95 %
1. Place the plastic plug into the syringe outlet port to prevent
contamination of the internal volume of the syringe. Each syringe
is shipped with a plastic plug in the outlet port.
2. Place the syringe assembly in a clean plastic bag and close
the end.
When maintained properly and recalibrated on a regular basis
(annual), SIBELMED calibration syringes should optimally function
within specifications. Visual inspection of all the components of
the assembly along with the functional tests as described will
identify a defective syringe.
Symbols
MANUFACTURER
(manufacture date, name and address of the manufacturer)
CE MARKING
MEDICAL DEVICE
CATALOGUE NUMBER
SERIAL NUMBER
TEMPERATURE LIMITS
Useful life
The useful life of the syringe is 5 years. When maintained
properly and recalibrated on a regular basis (annual), SIBELMED
calibration syringes should optimally function within specifications.
Visual inspection of all the components of the assembly along with
the functional tests as described will identify a defective syringe.