5) ADVERSE REACTIONS

To date, no adverse reactions have been reported.
6) STEP-BY-STEP INSTRUCTIONS
Adjusting the power of the ultrasonic generator.
N°1
N°2
N°3
N°4
N°5
Power settings for other ultrasonic units are available on Dentsply Sirona website.
Available on Dentsply Sirona website, www.dentsplysirona.com, a video illustrating the use of each
Start-X
insert. On request, this video can be mailed.
®
In general, to maximize clinical performance and safety, and to reduce the potential of instrument
breakage, start with lower power settings and only increase the power as required to accomplish the
clinical task. All tips should be used with minimal pressure and a light brushing motion.
Irrigation is recommended when performing procedures (such as dislodging metal posts – tip #4) that
require high energy to avoid overheating both the insert and the tooth.
7) DISINFECTION, CLEANING AND STERILIZATION
Reprocessing procedure for dental instruments
I - FOREWORD
Devices that are marked as "sterile" do not require any specific treatment before the first use.
For all other devices not labelled "Sterile", cleaning and sterilization prior first use is required
according to section III - STEP-BY-STEP INSTRUCTIONS part 4 to 8 of this DFU.
For those devices that are not labelled "single use", re-processing of the devices should be carried out
as per this DFU. For hygiene and sanitary safety purposes, these instruments must be cleaned and
sterilized before each re-use to prevent any contamination.
Excluded devices:
Uniclip and Mooser Calcinable plastic posts cannot be sterilized and must be disinfected by immersion
NaOCl (2,5 % at least) during 5 min. at ambient temperature.
B EN STAX DFU WEB / Rev.07 / 11-2017
SATELEC / P5 Newtron
Min
Max
16
16
10
18
7
Min
18
18
18
20
18
EMS miniPiezon
Max
8
9
8
9
4
7
9
10
3
8
2/6