SECTION 8—TROUBLESHOOTING GUIDE

Warnings and Alarms

NOTE: In all cases, pressing and holding the On/Off button for 2-3 seconds will
turn off and reset the unit.
LCD DISPLAY:
DESCRIPTION:
(Only possible with Pulse Mode)
Transportable oxygen concentrator has
not detected a breath for a predetermined
time period.
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens. This warning will be turned off for a minimum of 5
minutes.
LCD DISPLAY:
DESCRIPTION:
(Only possible with Pulse Mode)
Your breathing rate has begun to exceed
the capacity of the transportable oxygen
concentrator.
ACKNOWLEDGEMENT REQUIRED:
A sufficient reduction of your breath rate will automatically turn off the warning.
Alternately, press Return / Highlight button. Display will either show additional
warnings / alarms or revert to normal operating screens. This warning will be turned off
for a minimum of 5 minutes.
Invacare SOLO
2
INDICATORS:
Triple audible beep every 10 sec
YELLOW indicator Flashing.
SOLUTIONS:
1. Verify the cannula is connected, not
kinked, properly positioned and you are
breathing through your nose.
INDICATORS:
Single audible beep every 15 sec
YELLOW indicator On.
SOLUTIONS:
1. Immediately reduce your activity level to
get your breath rate to slow down.
2. Temporarily switch to Continuous Flow
Mode.
36
Part No 1163190