Indicators, Warnings, And Alarms; Filters - Invacare XPO2 100 Manuel D'utilisation

Concentrateur d'oxygène portatif
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1.3.2 Indicators, Warnings, and Alarms

Start-Up Indicator
Breath Detect Indicator
Internal Battery Charging Indicator
Audible Alarms
For all alarms and warnings, refer to 8.1 Troubleshooting,
page 33.

1.3.3 Filters

Air enters the XPO
concentrator through an air intake filter located
2
under the cover on the front of the unit. This filter prevents hair and
other large particles in the air from entering the unit. Before you
operate the XPO
concentrator, make sure this filter is clean, dry
2
and properly positioned.
1148112-F~12
The XPO
concentrator will turn on all lights briefly and check for stuck buttons upon power up. It will
2
then output a brief audible beep when first turned on to indicate the unit is operational.
The XPO
concentrator will deliver a measured pulse of oxygen each time it detects an inhalation. The
2
blue flow setting indicator light will blink each time an inhalation is detected and the XPO
outputs a pulse of oxygen.
When the concentrator is not turned on, but the AC, or DC, charger is connected to the concentrator
and providing power, the Battery Gauge Display will sequentially illuminate each of the four segments.
Periodically the sequential illumination will be interrupted by the actual battery charge display.
If the concentrator is fully charged, all four segments will be illuminated continuously until the
concentrator is turned on, or removed from charger power. If no Battery Gauge Display segments
are illuminated, the concentrator is not charging. The lack of charging could be due to either a loss of
power, loose charger connections or the internal battery is not at the proper temperature for charging.
Allow the unit to warm, or cool, to recommended charging temperature range (see 2.1 Typical Product
Parameters, page 9 ), check for loose connections and/or try a different power outlet.
To temporarily silence the audible alarms, quickly press the On/Off button. Except for the system Too
Hot/Too Cold alarms, this feature will reset itself, every 30 seconds, each time a new alarm is triggered,
or if the unit is turned Off. When performed during the system Too Hot/Too Cold alarms, the audible
alarm will be silenced until a new alarm is triggered, or the concentrator is turned Off.
To clean/replace the air intake filter, refer to 6.2 Cleaning the Air
Intake Filter, page 29.
General
concentrator
2
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Xpo2 100b

Table des Matières