7 Maintenance
7.1 Maintenance and servicing
If the following cleaning and safety notices are observed, the product
will be maintenance free.
IMPORTANT!
The hand control may only be opened by the
manufacturer of the bath lift.
1.
Check the product regularly for damage and to ensure that it is
securely assembled.
2.
In the event of damage and/or necessary repairs, contact the
dealer immediately.
3.
The bath lift must be checked by an authorized dealer before
every re-use or after 24 months at the latest (see Manufacturer's
recommendation for rehab products of the SPECTARIS industrial
association).
4.
If the bath lift is not used for a prolonged period, charge the
battery completely and then raise and lower the bath lift 3 times
without load. Repeat this process every 12 months. This helps
to maintain the service life of the battery.
7.2 Cleaning and disinfection
The product can be cleaned and disinfected using commercially
available agents.
1.
Disinfect the product by wiping down all generally accessible
surfaces with disinfectant (after disassembling the product, if
possible).
1471194-A
Cleaning the seat and backrest
CAUTION!
Danger of pinching fingers
– Be especially careful when cleaning in the area of the
lifting scissors, as they may move and cause injury.
IMPORTANT!
Risk of damage
– Never use a high-pressure cleaner to clean the
backrest.
1.
Wash the product using a cloth or a brush.
2.
Rinse the product with warm water.
3.
Dry the product with a cloth.
Lightly grease the guide for the sliding shoe (Fig. Sec. 3.1, H)
4.
using Vaseline.
Cleaning the covers (optional)
1.
Unbutton the covers from the seat and backrest.
2.
Wash the covers in a washing machine at max. 60 °C using
a mild detergent.
IMPORTANT!
Risk of damage
– Washing at high temperatures can cause shrinkage.
– Do not dry covers in the tumble dryer.
Maintenance