Technical Data - Seca 336 Mode D'emploi

Masquer les pouces Voir aussi pour 336:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
Er:012:
...
appears in the display?
– The scale was switched on with too great a load on it. Remove load and restart.
8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal
requirements. The year of initial calibration can be found behind the CE symbol on the
rating plate above the number of the appointed office, 0109 (Hessen Weights & Measures
Office). Recalibration is essential if one or more security marks are damaged or the
calibration counter does not match the number on the applicable calibration counter mark
(cf. page 15).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.

9. Technical data

Dimensions
length:
width:
height:
Weight
Temperature range
Maximum load
Graduations
Calibration as per Directive
2009/23/EC
Accuracy at first calibration
up to 5 kg:
5 kg up to 15 kg:
Medical product
as per Directive
93/42/EEC
Tare range
Height of figures
Power supply
Power consumption
Model 336
300 mm
640 mm
105 mm
approx. 3.5 kg
+10 °C to +40 °C
15 kg
10 g
Class III
± 5 g
± 10 g
Class I
up to 10 kg / 22 lbs
20 mm
Battery/
(option mains unit)
approx. 25 mA
17
GB

Publicité

Table des Matières
loading

Table des Matières