Minebea Intec Signum 2 Notice D'installation page 14

Masquer les pouces Voir aussi pour Signum 2:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Example 3:
Entering/modifying values, in the weights units defined under menu
item 1.7.x, for a single-range scale in "Verified" configuration mode.
ADC-CON
Select the ADC-CON menu item
)
Confirm the ADC-CON menu item
(press ) k p
as needed)
Select the VERIF. menu item
Verif.
The "Verified configuration mode"
menu item is shown
)
Confirm the VERIF. menu item
Class
The "Verification class" menu item
is shown
(press ) ) (
as needed)
Confirm "Accuracy class 3"
or "Accuracy class 4"
Ranges
The RANGES menu item is shown
)
Confirm the RANGES menu item
(press k repeatedly
if necessary)
Select the SINGLE menu item
SINGle
The "Single range scale" menu item
is shown
(
Confirm the SINGLE menu item
Weights
The WEIGHTS menu item is shown
)
Confirm the WEIGHTS menu item
The "Verification scale interval" menu
E
item (E) is shown
)
Confirm the E menu item
( ) k p
Enter the desired value (e.g., 0.002 kg)
(press ) repeatedly
if necessary)
Display the E menu item again
k
Select the MAX. menu item
Follow this procedure to enter values for:
Maximum capacity (e.g., 30 kg)
max.
Continue as described under Example 1,
after the step for entering the
maximum capacity
14
Signum Option A15
Example 4:
Entering/modifying values, in the weights units defined under menu
item 1.7.x, for a multi-interval scale in "Verified" configuration mode.
(The procedure for configuring a multiple range scale is similar.)
ADC-CON
Select the ADC-CON menu item
)
Confirm the ADC-CON menu item
(press ) k p
as needed)
Select the VERIF. menu item
Verif.
The "Verified configuration mode"
menu item is shown
)
Confirm the VERIF. menu item
Class
The "Verification class" menu item
is shown
(press ) ) (
as needed)
Confirm "Accuracy class 3/4"
Ranges
The RANGES menu item is shown
)
Confirm the RANGES menu item
(press k repeatedly
if necessary)
Select the MULT.INT. menu item
Mult.Int
The "Multi-interval scale" menu item is
shown
(
Confirm the MULT.INT. menu item
Weights
The WEIGHTS menu item is shown
)
Confirm the WEIGHTS menu item
The "Verification scale interval" menu
E
item (E) is shown
)
Confirm the E menu item
( ) k p
Enter the desired value (e.g., 0.002 kg)
(press ) repeatedly
if necessary)
Display the E menu item again
k
Select the RANGE 1 menu item
Follow this procedure to enter values for:
Range 1
Limit for range 1 (e.g., 6 kg)
Limit for range 2 (e.g., 15 kg)
Maximum capacity (e.g., 30 kg)
Continue as described under Example 1,
after the step for entering the maximum
capacity

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Signum 3

Table des Matières