Example 1:
Entering/modifying values, in the weights units defined under menu
item 1.7.x, for a single-range scale in "Standard" configuration mode.
ADC-CON
Select the ADC-CON menu item
)
Confirm the ADC-CON menu item
(press k as needed)
Select the STAND. menu item
Stand.
Standard configuration mode
)
Confirm the STAND. menu item
The RANGES menu item is shown
Ranges
)
Confirm the RANGES menu item
(press k repeatedly
if necessary)
Select the SINGLE menu item
SINGle
The "Single range scale" menu item
is shown
(
Confirm the SINGLE menu item
Weights
The WEIGHTS menu item is shown
)
Confirm the WEIGHTS menu item
The "Scale interval" menu item (D)
d
is shown
)
Confirm the D menu item
( ) k p
Enter the desired value (e.g., 0.002 kg)
(press ) repeatedly
if necessary)
Display the D menu item again
k
Select the MAX. menu item
max.
The "Maximum capacity" menu item
is shown
)
Confirm the MAX. menu item
( ) k p
Enter the desired value (e.g., 30 kg)
(press ) repeatedly
if necessary)
Display the MAX. menu item again
(
The UNITS menu item is shown
(press ) as needed)
Select the "Available weight units"
menu item (UNITS)
k
The SAVE menu item is shown
Save
) k )
Save the values entered (YES)
or quit without saving values (NO)
Example 2:
Entering/modifying values, in the weights units defined under menu
item 1.7.x, for a multi-interval scale in "Standard" configuration mode.
(The procedure for configuring a multiple range scale is similar.)
ADC-CON
Select the ADC-CON menu item
)
Confirm the ADC-CON menu item
(press ) k p
as needed)
Select the STAND. menu item
Stand.
Standard configuration mode
)
Confirm the STAND. menu item
Ranges
The RANGES menu item is shown
)
Confirm the RANGES menu item
(press k repeatedly
if necessary)
Select the MULT.INT. menu item
Mult.Int
The "Multi-interval scale" menu item
is shown
(
Confirm the MULT.INT. menu item
Weights
The WEIGHTS menu item is shown
)
Confirm the WEIGHTS menu item
d
The "Scale interval" menu item (D) is
shown
)
Confirm the D menu item
( ) k p
Enter the desired value (e.g., 0.002 kg)
(press ) repeatedly
if necessary)
Display the D menu item again
k
Select the RANGE 1 menu item.
Follow this procedure to enter
values for:
Limit for range 1 (e.g., 6 kg)
Range 1
Limit for range 2 (e.g., 15 kg)
Range 2
Maximum capacity (e.g., 30 kg)
max.
Continue as described under
Example 1, after the step for entering
the maximum capacity
13