Calibration Of Axes - Amann Girrbach ceramill map 200+ Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

I N S T A L L A T I O N
5.7

Calibration of axes

Following the set-up and connection to the PC, the
E N
desktop scanner must be calibrated using the sup-
plied calibration model.
The scanner should be recalibrated at reg-
ular intervals.
The device must be calibrated each time
after transporting.
▷ Place calibration model on the scan base plate.
Fig. 5
Calibration model in scanner
▷ Create a new patient case in the data base and
start the scanner software from the data base.
▷ Cancel the subsequent command prompt for
the 2D scan.
▷ Click on
Scanner Options > Axis calibration
start the axis calibration.
Fig. 6
26
In the subsequent window, the values #1 and #2
must correspond with those of the calibration
model. If the values do not correspond, see
Section 5.7.1.
▷ Start the axis calibration.
Fig. 7
5.7.1
Entering calibration object values
▷ To enter the calibration object values, click on
Application... > Register Calibration Object...
Fig. 8
▷ Enter calibration object values and confirm with
OK
.
to
Fig. 9
▷ After entering the values, a calibration process
must be carried out.
.

Publicité

Table des Matières
loading

Table des Matières