Correct And Proper Use; Information In The Event Of An Emergency - ProMinent DULCOTEST CLO 1 Guide D'utilisation

Sonde pour le chlore libre
Masquer les pouces Voir aussi pour DULCOTEST CLO 1:

Publicité

Les langues disponibles

Les langues disponibles

CAUTION!
Prerequisites for operation
Possible consequence: Slight or
minor injuries, material damage
The sensor may only be used in
flow gauges that ensure the cor‐
rect flow parameters.
There must be a free outlet or at
most a counter pressure of 1 bar
at the outlet of the flow gauge.
The maximum operating pressure
of the respective single compo‐
nents must be observed.
The sensor's voltage supply may
not be interrupted
Following longer interruptions to
the voltage supply (> 2 h), allow
the sensor to run-in again and
recalibrate it
1.4

Correct and Proper Use

NOTICE!

Correct and Proper Use

The sensor may only be used to
determine and regulate concen‐
trations of free chlorine
The sensor may not be used in
conjunction with organic chlorine
compounds, such as trichloroiso‐
cyanuric acid, or stabilisers, such
as cyanuric acid
Any other uses or modifications
to the system are prohibited
The sensor is not a safety com‐
ponent
1.5
Information in the event of
an emergency
In the event of an emergency, switch
n
off the controller
If liquid exits from the flow gauge
n
housing then close the shut-off valves
provided by the customer at their
inlets and outlets
Observe the safety instructions of the
n
plant operator before opening the flow
gauge housing
In case of proven over-metering,
n
destroy the excess chlorine using a
suitable reducing agent (e.g.
hydrogen peroxide or sodium sul‐
phite)
Introduction
13