Assembly
4.
Fill the sensor cap (4) to the brim
with electrolyte (3) taking care to
avoid bubbles
Fig. 3: Sensor cap (4) with electrolyte (3)
5.
Place the sensor shaft (5) vertically
onto the sensor cap (7) filled with
electrolyte (6) and turn until the
thread catches
Fig. 4: Sensor shaft (5) with electrolyte (6)
and sensor cap (7)
6.
Turn the sensor cap manually until
it stops. No gap must be visible
between the sensor cap and the
sensor shaft
20
3
4
Fig. 5: Electrolyte (8) exits
A0108
7.
5
6
7
A0126
ð Excess electrolyte (8) exits
from the openings when
screwing together.
Remove sensor cap protector (11)
8
A0127