.. Device connection, cables at side (Option)
• Remove cable cover (Figure 48)
• Expose USB and supply cables (left or right aperture on rail
cover)
• With a knife, cut towards the snap-off point provided on the left
or right of the cable cover (Figure 49)
• Carefully use fingers to snap off the cut part (Figure 50)
• Insert both cables in the appropriate box and fix with the tension
compensator.
• Join both cables together with spiral cable and attach firmly using
the cable channel on the rail cover (Figure 51)
• Replace cable cover
. Computer connection
• Connect the LS 900
to the PC with the enclosed USB2.0 cable.
®
Use of the 2 m cable supplied with delivery is recommended.
BEWARE
The USB cable should not exceed 5 m in length,
otherwise impairment of device function must
be expected.
• Insert mains cable. Integral mains components work with the vol-
tages specified under section 7.1 'Technical data'. It is not necessary
to select the voltage on the device.
• If an instrument table has been supplied, the power supply of the
LS 900
can beconnected to the Switchbox SB01 (left drawer). Use the
®
manual enclosed to the switchbox and the instrument table.
.. Software
Install software in accordance with separately described 'Software
Installation'.
48
49
50
51
1500.7220055.04030
© Haag-Streit AG, CH-3098 Koeniz, Switzerland