• the systolic or diastolic pressure was outside the measure-
ment range ("Hi" or "Lo" appears on the display)
• the cuff is fastened too tightly or too loosely
(
or
appears on the display)
• the blood pressure is higher than 297 mmHg for more than
1.5 seconds (
appears on the display)
• pumping up takes longer than 180 seconds (
the display)
• there is a system or unit error (
on the display)
• the batteries are almost empty

8. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp cloth
only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects on
the device and cuff. Remove the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicat-
ing either "     " or " Lo  " . In such cases, you should seek medical
assistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
appears on
,
,
or
appears
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.

9. Technical specifications

Model No.
Measurement
method
Measurement range
Display accuracy
Measurement inac-
curacy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
conditions
Power supply
23
BC 28
Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Cuff pressure 0 – 300 mmHg,
Systolic 60-260 mmHg,
Diastolic 40-199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
2 x 60 memory spaces
L 83.8 mm x W 59.9 mm x H 29.4 mm
Approximately 92 g (without batteries)
140 to 195 mm
+ 10°C to +40°C, ≤85% relative humid-
ity (non-condensing)
-20°C to +50°C, ≤ 85% relative humid-
ity, 700 –1060 hPa ambient pressure
2 x 1.5V
AAA batteries