The "HealthCoach" app must be active to allow data transfer.
If your smartphone has a protective cover, remove this to ensure
that there is no interference during the transfer. Begin the data
transfer in the "HealthCoach" app.
8. Cleaning and storing the device and cuff
• Clean the device and cuff carefully using a slightly damp cloth
only.
• Do not use any cleaning agents or solvents.
• Under no circumstances should you hold the unit under water,
as this can cause liquid to enter and damage the unit.
• If you store the device, no heavy objects should be placed on
top of it. The cuff line should not be bent sharply.
In such cases, repeat the measurement and/or data transfer.
Ensure that the cuff line is correctly attached and that you do not
move or speak during the measurement.
9. Error messages/troubleshooting
In the event of errors, the error message
display.
Error messages may appear if:
• it was not possible to record the pulse:
• you move or speak during the measurement:
• the cuff is fastened too tightly or loosely:
• errors occur during the measurement:
• the pump pressure is higher than 300 mmHg:
• the batteries are almost empty
•
the data could not be sent via Bluetooth
_ appears on the
1;
2;
3;
4;
5;
6;
:
.
®
30
The word mark Bluetooth
®
trademarks of Bluetooth SIG, Inc. Any and all use of these marks
by Hans Dinslage GmbH is done so under licence. Other trade-
marks and trade names are the property of the relevant holder.
10. Technical specifications
Model No.
SBM 37
Measurement me-
Oscillometric, non-invasive blood pres-
thod
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 250 mmHg,
Systolic 50 – 250 mmHg,
Diastolic 30 – 200 mmHg,
Pulse 40 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 139 mm x W 94 mm x H 48 mm
Weight
236 g (without cuff, without batteries)
Cuff size
22 to 36 cm
Permissible operating
+ 5°C to + 40°C, 15-93% relative humi-
conditions
dity (non-condensing)
and accompanying logo are registered