Medela Clario Toni Mode D'emploi page 327

Masquer les pouces Voir aussi pour Clario Toni:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 66
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, 3rd Edition, Table 2)
Electromagnetic Immunity
Clario Toni is intended for use in the electromagnetic environment specified below�
The customer or the user of Clario Toni should assure that they are used in such
environment�
Immunity Tests
IEC 60601
test level
Electrostatic
± 8 kV contact
Discharge (ESD)
± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV
transient/burst
for power
IEC 61000-4-4
supply lines
± 1 kV
for input /
output lines
Surge
± 1 kV
IEC 61000-4-5
line(s) to line(s)
± 2 kV
line(s) to earth
0 % U
Voltage dips,
short interrup-
for 0�5 cycle at
tions and voltage
0°, 45°, 90°,
variations on
135°, 180°, 225°,
power supply
270° and 315°
input lines
0 % U
IEC 61000-4-11
for 1 cycle
70 % U
for 25 cycles
at 50 Hz
Single phase:
at 0°
0 % U
for 250 cycles
at 50 Hz
30 A/m
Power frequency
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a�c� mains voltage prior to application of test level�
T
Compliance
level
± 8 kV contact
± 15 kV air
± 2 kV
for power
supply lines
± 1 kV
for input /
output lines
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth not
applicable
0 % U
T
T
for 0�5 cycle at
0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
T
T
for 1 cycle
70 % U
T
T
for 25 cycles
at 50 Hz
Single phase:
at 0°
0 % U
T
T
for 250 cycles
at 50 Hz
30 A/m
Electromagnetic
environment – guidance
Floors should be wood, concrete or
ceramic tile� If floors are covered with
synthetic material, the relative humidity
should be at least 30 %�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be that
of a typical commercial or hospital
environment� If the user of Clario Toni
requires continued operation during
power mains interruptions, it is recom-
mended that Clario Toni be powered
from an uninterruptible power supply
or a battery�
Power frequency magnetic fields
should be at levels of a typical
commercial or hospital environment�
Technical documentation
| 327

Publicité

Table des Matières
loading

Table des Matières