Power Supply And Battery Operation; Battery Low; Battery Empty - Medela Clario Toni Mode D'emploi

Masquer les pouces Voir aussi pour Clario Toni:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 66

2 Power supply and battery operation

The Clario Toni is a mains- and battery powered suction pump� Before you plug in the device,
please check that your local power supply is the same as the voltage given on the specifica-
tion plate� Used as described in this user guide, the Clario Toni AC / DC model has a maxi-
mum of 50 minutes of battery time before it is necessary to recharge� If the Clario Toni AC / DC
model is not in use, the battery must be calibrated every 60 – 90 days�

Battery low

During battery operation a slow blinking of the LED indicator indicates, that the rechargeable
battery is soon discharged� The function of the Clario Toni is maintained for a short time
(~10 minutes), but the battery must be recharged as soon as possible�

Battery empty

As soon as the rechargeable battery is empty, the pump comes to a stop� The proper function
of the pump is no longer maintained� The rechargeable battery must be recharged or replaced�
8 |
Power supply and battery operation

Publicité

Table des Matières
loading

Table des Matières