17
Technical Information
Technical Specification
Inlet flange
Foreline flange
Rotational speed
Start-up time
Recommended forepump
Operating position
Operating ambient temperature
Relative humidity of air
Bakeout temperature
Lubricant
Cooling requirements
Coolant water
Noise Pressure level (pump only)
Pwr supply:
Input Voltage:
Input Frequency:
Max input pwr:
Stand-by pwr:
Max oper. power:
Installation category
Pollution degree
Storage temperature
Max operating altitude
†
average value ± standard deviation.
296/352
CFF 4.5" O.D.
CFF 2.75" O.D.
KF16 NW
70000 rpm (1167 Hz driving frequency)
< 2 minutes
mechanical: Agilent DS 40M
mechanical: Agilent DS 102
dry pump: Agilent IDP-3
dry pump: Agilent IDP-7
Any
+5 °C to +35 °C
0 – 90 % (not condensing)
80 °C at inlet flange (ISO63) and 120°C (CFF 4.5") max
Permanent lubrication
Max forced cooling air temperature = 35°C
Max cooling water temperature = 25°C
Minimum flow: 65 l/h (0.30 GPM)
Temperature: +15 °C to +25 °C
Pressure: 2 to 4 bar (30 to 60 psi)
40 ± 4 dB(A) at 1 meter
76 Vac, 3 Phase (provided by the controller)
1167 Hz max (provided by the controller)
130 VA
8 W
70 W (according to controller model) with water or air cooling
II
2
-40° C to +70° C
3000 m
TwisTorr 74 FS User Manual / 87-901-053-01 (A.00)
(Min = 32 dB(A), Max = 48 dB(A))
†
ISO 63
KF 40