5 Operation
36
Operating Instructions Midwest
5

Operation

NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
CAUTION! A loose or partially removed bur instrument can come
loose from the head or break off. Risk of injury! Therefore the instrument
must only be used when the bur instrument is securely clamped at least
10 mm deep.
CAUTION! The handpiece must only be operated with the clamping
system closed.
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
®
®
E Pro / Midwest
E Plus / Midwest
66 04 206 D3686.201.01.04.02
®
E mini