Error Codes; Configuration Of The Serial Rs232C Interface - IKA IKAMAG RET control visc Mode D'emploi

L'agitateur magnétique, appareil de chauffage possédant une fonction d'agitatione
Masquer les pouces Voir aussi pour IKAMAG RET control visc:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error codes

Example of an error code
Er 1
Measuring sensor Pt100 not
inserted in A -mode of ope-
ration; LED „Temp" blinks
alternately green - red
Er 3
Internal temperature of instru-
ment exceeds 80°C LED „Temp"
blinks alternately red - orange
Er 4
speed deviation from rated to real
speed higher than ±300 1/min
LED „Mot" blinks green
Er 5
measuring sensor Pt100 not in me-
dium or wrong measuring sensor
inserted; LED „Temp" blinks red
Er 21 faulty safety relay
LED „Temp" blinks red
Er 22 firmly set safety circuit faulty;
LED „Temp" blinks red
Er 23 adjustable safety circuit faulty;
LED „Temp" blinks red
For setting back the error code
and newkalibrate the safety circuit
on a new safety temperature.
22
(see chapter „setting the safety
temperature")
Er 24 breakage of heating plate
thermocouple;
LED „Temp" blinks red
Er 25 polarity ot thermocouple;
LED „Temp" blinks red
heating is
switched off
Configuration of the Serial
heating is
switched off
- The functions of the interface lines between laboratory instrument
and automation system are selected from the specified signals of
heating and motor
the EIA-standard RS 232 C, corresponding with DIN 66 020 Part 1.
The allotment of the bushing can be taken from illustration 2.
are switched off
- For the electrical properties of the interface lines and for the allot-
ment of the signal status, standard RS 232 C, corresponding with
heating is switched
DIN 66 259 Part 1 applies.
off
- Transmission method:
heating is switched
- Mode of transmission:
off
- Character format:
heating is switched
off
heating is switched
off
- Transmission speed:
- Data flow control:
heating is switched
off
heating is switched
off
RS 232 C Interface
Asynchronous signal transmissi
on in start-stop-operation.
Fully Duplex.
Character imaging acc. to data
format DIN 66 022 for
start-stop-operation.
1 start bit
7 character bits
1 parity bit (even])
1 stop bit
9600 bit/s
Hardware handshake RTS/CTS
RET cv 0498

Publicité

Table des Matières
loading