The handpiece is intended for use in the electromagnetic environment specified below. The Customer or the user of the handpiece should assure
that it is used in such an environment.
Immunity test
IEC 60601
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD) IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1KV for input/
output lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common mode
Voltage dips, short
<5 % UT for 0,5
interruptions and
cycle
voltage variations on
(>95 % dip in UT )
power supply input
lines IEC 61000-4-11
40 % UT for 5 cycles
(60 % dip in UT )
70 % UT or 25 cycles
(30 % dip in UT )
<5 % UT for 5 sec
(>95 % dip in UT )
The handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the handpiece should assure
that it is used in such an environment.
Immunity test
IEC 60601
Conducted RF
3Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
BENENACDFUWEB / Rev.07 / 09-2019
Compliance
test level
level
±2, ±4, ±6KV Contact
±2, ±4, ±8KV Air
Not applicable
Not applicable
Not applicable
Compliance
test level
level
3 V
3 V/m
Electromagnetic environment –guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the handpiece requires continued operation
during power mains interruptions, it is recommended that the handpiece
be powered from an uninterruptible power supply or a battery.
Electromagnetic environment –guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the handpiece, including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
should be less than the compliance level in
6/8