6) STEP BY STEP INSTRUCTIONS
1) Prepare canal to produce a fully tapered shape.
2) Fill pulp chamber with NaOCl, EDTA, or other final rinse solution.
3) Select the Activator tip that manually fits loosely within 2 mm of working length.
4) Place the barrier sleeve over the complete length of the handpiece.
5) Attach the Activator tip over the barrier-protected handpiece. The Activator should snap on firmly,
promoting a secure connection with the handpiece.
6) Place the attached Activator tip into the prepared root canal.
7) Depress the ON/OFF switch to activate. Note: Switch defaults to high speed upon activation. Depress the
3-speed switch to select medium speed or low speed.
8) Use a pumping action to move the handpiece/Activator in short 2-3 mm vertical strokes.
9) Hydrodynamically agitate the intracanal solution for 30-60 seconds.
10) Irrigate, then use intracanal suction to eliminate loose debris.
11) Repeat the above steps for each intracanal irrigant used.
When the clinical procedure has been completed, remove the attached tip by grasping the large circular
clean guard portion of the connected Activator with fingers and snap off. Pull the Activator off the
handpiece by firmly supporting the contra-angled neck of the handpiece. Next, remove the barrier sleeve
and discard. Activator tips and protective barrier sleeves are intended for single-patient use only.
6.1) ACTIVATOR TIP REMOVAL
BENENACDFUWEB / Rev.07 / 09-2019
Firmly support the head of the handpiece with your thumb while grasping
the white, circular, flange portion of the Activator tip with opposite thumb and
forefinger.
Use your thumb to maintain support of the handpiece and pull the Activator tip
STRAIGHT off. Keep the handpiece head and Activator tip aligned during the
removal process.
The Activator tip snaps ON/OFF in exactly the same alignment. NEVER use
a clockwise or counterclockwise motion to remove a Tip. NEVER twist, turn or
torque the tip when removing.