8 Adjustment – Easy Calibration

8.2 Restore factory setting

a.
1 x
> 3 s
b.
c.
1 x
70
Operating manual
The CAL symbol permanently shown in the display indicates
that an adjustment has been made.
a.
Enter CAL mode: Pressing and holding (> 3 s) the menu
key invokes the CAL mode.
ð
The display shows 'on'.
ð
'CAL' flashes.
b.
Switch off CAL mode: Pressing one of the arrow keys de-
activates the CAL mode.
ð
The display changes from 'on' to 'off'.
ð
'CAL' continues to flash.
c.
Restore factory condition: Press 'Enter' key.
ð
The permanently displayed CAL icon has disappeared.
ð
The instrument has been returned to the original fac-
tory condition.
For the Transferpette® -8/-12 electronic , a volume offset is
applied during adjustment (i.e. the volume changes by the
same amount over the entire volume range of the pipette). It
is recommended to perform the adjustment at 50% of the
nominal volume.
The instrument is permanently adjusted for aqueous solu-
tions but can also be adjusted for solutions of different den-
sity, viscosity, and temperature. The Transferpette® -8/-12
electronic can be adjusted in any mode (except GEL mode).