Source language: German
Contents
Declaration of Conformity
Declaration of Conformity
We declare under our sole responsibility that the product to which this declaration
relates is in conformity with the Pressure Vessel Directive 2014/68/EU and the
regulations AD 2000.
Explanation of symbols
EN
This symbol identifies information that is absolutely essential for
Page
your safety and health.
11
Failure to observe this information may result in negative effects for
11
your health and injuries.
11
13
This symbol identifies information that is important for problem-free opera-
14
tion as well as for working with the decomposition vessel.
14
Failure to observe these instructions may result in inaccurate measure-
15
ment results or damage to the decomposition vessel.
15
15
15
Safety instructions
16
16
17
17
17
17
18
18
18
Operating requirements:
19
The maximum amount of energy input into the decomposition vessel must not
19
exceed 40000 J. (Select the weight of the sample accordingly). The permissible
19
operating pressure of 230 bar (23 MPa) must not be exceeded. The maxi-
94
mum permissible operating temperature must not exceed 50 °C.
95
Do not fill the decomposition vessel too full of the sample. Only fill the decompo-
sition vessel with oxygen up to a maximum pressure of 40 bar (4 MPa). Monitor
the adjusted pressure on the pressure reducer.
Perform a check before every combustion to ensure there are no leaks (see
sections "Leakage test").
Intended purpose:
The C 6010 and C 6012 decomposition vessels must only be used
for determining the gross calorific value of solid and liquid com-
bustible materials in accordance with DIN 51900, BS 1016 T5, ISO
1928, ASTM 5468, ASTM 5865, and ASTM 4809.
Operation of decomposition vessels is only permitted in combina-
tion with the IKA
calorimeter C 6000 g/iso.
®
11