Replacing the ignition electrode
Take care two new o-rings
(item 10 and 23) when
inserting the new electrode.
23
10
Item 6
Shaft
Item 10 O-ring
Item 13 Ignition electrode
Item 23 O-ring
replacing ignition wire
Item 31 Hex nut M4
Replacing the crucible holder
8
Loosen hex nut (item 30).
Item 8 Crucible holder
Item 30 Hex nut M3
30
replacing groundelectrode
13
6
31
replacing ignition electrode
replacing sealing ring
Replacing the ground electrode
Loosen hex nut (item 30).
Item 7
Ground electrode
Item 30
Hex nut M3
Maintaining oxygen valve
The oxygen valve is used to fill and empty the decomposition vessel. The o-ring
(item 22) may wear because of aggressive combustion gases, which means that
the decomposition vessel may not be tight.
The oxygen valve must be unscrewed to change the o-ring.
Q
16
15
21
22
17
20
Item 15 Screwing
Item 16 Valve piston
Mounted oxygen
Item 17 Nozzle
valve:
Item 20 Spring
Item 22 O-ring
replacing ignition wire
7
30
replacing groundelectrode
15
21
16
22
20
17
17
replacing ignition electro
replacing sealing rin