Controller Specifications
Tab. 4
Input:
Voltage
Max Power
Operating conditions: Max Altitude:
HV-OUTPUT:
Voltage
Ripple VPP MAX @ 3Kv
Current max
Maximum Output Power
Current Recorder Output signal
Operating Temperature
Storage Temperature
Compliance to Norms
Line fuse
High Voltage Connection
Ion Pumps Controller 24 V User Manual / 87-900-127-01 (C)
Technical Information
Controller Description
24 Vdc +/- 10 %
7 W
3000 m
Default Value = ± 3000 V ± 5 % (mean Value)
300 V
4 mA ± 10 %
4 W
0 to 10 V linear 1 µA = 10 mV and 1 mA = 10 V
0 °C to 45 °C
-10 °C to 75 °C
EN 61010
EN 61326-1
T1A slow
1 High Voltage Fischer connectors
5
33/42