Tab. 5
Recorder Output Risolution:
Recorder Output Sensibility:
Recorder Output Rload >
Recorder Output Ripple max
Output Polarity Selection
The output polarity is selected by the high voltage multiplier card.
Polarity Change Procedure
1.
2.
3.
4.
5.
Ion Pumps Controller 24 V User Manual / 87-900-127-01 (C)
pp
Switch off the power and unplug the power cord from the
controller rear panel socket.
Undo the 4 screws and then remove the cover
Disconnect the H.V. white wires (1) from the multiplier board,
remove the board and rotate it 180 ° horizontally; reinstall the
board into the socket and reconnect the H.V. white wires.
Install the cover and screws.
Connect the power cord then switch on the unit and check the
output polarity on the front panel indication.
1 µA
1 µA
100 kΩ
<
10 mVpp
Technical Information
Controller Description
5